A CMOS Electrochemical Biochip With 32× 32 Three-Electrode Voltammetry Pixels
Arun Manickam, Robert G. Kuimelis, Arjang Hassibi, Kae-Dyi You, Nicholas A. Wood, Lei Pei, Yang Liu, Rituraj Singh, Nader Gamini, Mark W. McDermott, Davood Shahrjerdi
IEEE Journal of Solid-State Circuits 54 (11), 2980-2990 (2019)
Abstract
A CMOS-integrated electrochemical biochip for high-performance molecular testing is presented. The system includes an array of 32×32 three-electrode voltammetry pixels and on-chip temperature control between 25 and 95 °C. Each 100 μm × 100 μm pixel includes a CMOS-compatible and chemo-stable amorphous carbon electrode transducer connected to in-pixel ΣA current detection circuitry with 280 fA rms input-referred noise (0.1-20 Hz bandwidth) and 93 dB dynamic range (DR). Array-based DNA detection assays are implemented, and successful DNA melt-analysis and real-time label-free DNA hybridization detection are reported.